Reprocessing
13.7
Surface disinfection with cleaning
13.7.1 Surface
disinfection
of the device surfaces
1. Remove soiling immediately. Use a cloth dampened with disinfectant to remove
soiling.
2.
WARNING
Risk due to penetrating liquid
Penetrating liquid may cause the following:
– Damage to the device
– Electric shock
– Device malfunctions
►
Ensure that no liquid penetrates the device.
CAUTION
Risk of contaminating the device
If the water trap is removed, fluid may get into the main device while it is being
disinfected.
►
Leave the water trap fitted during surface disinfection.
Perform the surface disinfection of the entire device by wiping backwards and
forwards at least 3 times:
– In particular, disinfect device surfaces and hand contact points that are
touched frequently.
– Follow the list of validated disinfectants (see "Validated reprocessing
The following illustration shows the frequently touched surfaces and hand
contact points:
39472
No.
Designation
1
Screen
2
Control dials of the vaporizers