102
Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Operation
7.4.3 CBM
mode
The CBM mode allows patient monitoring without unnecessary alarms during
extracorporeal oxygenation of the patient by a heart-lung machine.
Properties of CBM mode:
– All gas concentrations are measured independently of the respiratory phases.
– The CO
2
apnea alarms and pressure apnea alarms are inactive.
The CBM mode can be used in all active ventilation modes.
When ventilation modes are changed, the CBM mode remains active. Changing to
the standby mode deactivates the CBM mode.
Deactivating the CBM mode activates the apnea monitoring.
Activating
1. Open the
Alarms
dialog.
2. Touch the
Settings
tab (1).
3. For
Cardiac bypass mode (CBM)
, touch the
On
button (3).
Deactivating
Deactivate the CBM mode optionally as follows:
►
For
Cardiac bypass mode (CBM)
, touch the
Off
button (2):
►
In the main menu bar, touch the
Exit CBM
button.
35307
2
3
1
Alarms