Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Overview
Electronically controlled
gas mixer (model desig-
nation A350)
– The fresh-gas delivery is adjusted using the screen
with the aid of settable parameters for O
2
concen-
tration in % and for fresh-gas flow in L/min.
– Depending on the version, Air or N
2
O can be
selected as the carrier gas.
– With O
2
flowmeter for O
2
insufflation
– With integrated, mechanical emergency O
2
deliv-
ery
Add. O
2
External fresh-gas outlet
– Allows use with external non-rebreathing systems,
e.g.:
– Mapleson
– Kuhn
– Bain
– Magill
– Waters
Advanced Cylinder Sup-
port
On devices that are equipped with Advanced Cylinder
Support, the gas cylinder valves can also remain open
during operation with the central supply.
Integrated O
2
monitoring
– Measurement, monitoring, and display of the inspi-
ratory O
2
concentration
– The measurement is performed by the O
2
sensor
integrated in the breathing system.
Integrated patient-gas
measurement module
– Measurement, monitoring, and display of the inspi-
ratory and expiratory gas concentrations of O
2
,
anesthetic gases, CO
2
, and N
2
O
– Detection and indication of anesthetic gas mixtures
– Display of the xMAC
O
2
detection
– During the self-test, a check is made using the inte-
grated patient-gas measurement module to ensure
that the connected O
2
supply is actually delivering
O
2
.
Support of Infinity ID
accessories
Allows the use of Dräger Infinity ID accessories with
the following functions:
– Generates a message when the maximum period
of use is exceeded for the breathing circuit, the
water traps, the CO
2
absorber, and the flow sensor
– Generates a notice when the breathing circuit is
incorrectly connected
– Generates a message if the CO
2
absorber is not
present or is not locked
Active anesthetic gas
scavenging
Anesthetic gas scavenging with flow indicator for use
with an active disposal system with a wall terminal unit
Passive anesthetic gas
scavenging
Anesthetic gas scavenging for gas disposal without an
active disposal system
Name
Description