Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
125
Operation
7.14.2
Checking or replacing consumables
Prerequisite: The device is in standby mode.
Integrated patient-gas measurement module
1.
WARNING
Risk of infection
The water trap may contain infectious fluid.
►
Proceed carefully when emptying and take protective measures if necessary.
►
Follow the infection prevention policies and reprocessing regulations of the
healthcare facility.
Check the water trap of the patient-gas measurement module (PGM) for
leakage. If necessary, empty or replace the water trap.
2.
WARNING
Risk of infection
Used sample lines and water traps may be infectious due to the breathing gases
that passed through them.
►
Replace the sample line and the water trap regularly in the following
situations:
– If the sample line is connected to the filter on the Y-piece, replace it daily.
– If there is no filter fitted to the Y-piece and the sample line is connected
directly to the Y-piece, replace the sample line after every patient.
►
Remove the sample line from the water trap.
►
Initially leave the water trap fitted to prevent infectious fluid from spurting out.
Remove the water trap only after surface disinfection.
►
Replace the water trap after each patient.
If no filter was used, replace the sample line and dispose of the used sample
line.
Vaporizer filling level
►
Check the vaporizer filling level in the sight glass. Fill the vaporizer if required.
CO
2
absorber
►
Check the soda lime for discoloration and replace if necessary, see page 75.
Breathing hoses and filters
1. Replace the hoses and filters according to infection prevention policy of the
hospital.
2. Connect a suitable breathing circuit and filters, see page 67.
7.14.3 Checking
the
device
Prerequisite: The device is in standby mode.
1. Perform the leakage test, see page 127.
2. Flush the breathing system if necessary.