Operation
–
Auto
: This setting is available when an anesthetic gas monitor is connected
via the COM 2 serial port (e.g., Dräger Vamos). The flow measurements are
automatically corrected for the anesthetic gas detected by the anesthetic gas
monitor. If no anesthetic gas is detected or if the data connection to the
anesthetic gas monitor is interrupted, no compensation takes place and the
message
None
(4) is displayed.
7.3.5
Changing the ventilation mode
1. In the therapy bar, touch the tab (1) of the new ventilation mode.
When the ventilation mode is changed, the ventilation settings are adopted from
the previous ventilation mode or they are sensibly derived.
In addition, the alarm settings are adjusted to reasonable values, see page 159.
2. Adjust the therapy controls (3) or the buttons (2).
3. Activate the ventilation mode with the rotary knob. A signal tone is emitted when
the mode is changed.
7.3.6 Synchronizing
the
breaths
Prerequisite: The device has the "Spontaneous breathing support" option.
Turning on the synchronization activates the set pressure support, for example, see
page 298.
1. Turn the synchronization on or off with the buttons
SIMV
(1) or
CMV
(2).
2. If required, show the extended therapy bar with one of the
More
buttons (3) or
(4). Then adjust the additional parameters (
Trigger
,
∆
Psupp
, etc.).
WARNING
Risk of insufficient ventilation
In ventilation modes in which breaths are to be triggered only by the patient (e.g.
PSV
), adverse settings or sensor failure can lead to insufficient ventilation.
►
To maintain a minimal ventilation of the patient, set the respiratory rate to a
suitable value.
35268
1
3
2
3
2
35279
3
4
2
1