![Dräger Atlan A300 Скачать руководство пользователя страница 162](http://html.mh-extra.com/html/drager/atlan-a300/atlan-a300_instructions-for-use-manual_2526728162.webp)
162
Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Alarms
Gas measurement alarms with inspiratory O
2
measurement
Ventilation alarms
Alarm
Priority
Priority
Priority
Note
(Low)
Caution
(Medium)
Warning
(High)
FiO
2
low
---
After at least 5 seconds
FiO
2
high
---
After at least 5 seconds
O
2
measurement not
available
O
2
measurement tem-
porarily inaccurate
O
2
sensor not ready
>20 s
---
---
Alarm
Priority
Priority
Priority
Note
(Low)
Caution
(Medium)
Warning
(High)
Apnea (no flow)
Apnea (no pressure)
Apnea
---
At the latest after
20 seconds
(for RR
≥
6)
or
At the latest after
35 seconds
(for RR <6)
or
at the latest after
65 seconds for the
Pause, Manual / Sponta-
neous and Ext. FGO
modes
15 seconds later
(for RR
≥
6)
or
30 seconds later
(for RR <6)
Apnea Ventilation
At the latest after 20 seconds
(15 seconds for RRmin
≥
4)
(configurable, see page 168)
---
Tidal volume not
achieved
---
After 3 consecutive
breaths
---
Airway press. continu-
ously high
---
---
>15 seconds above the
manually or automati-
cally set limit
Airway pressure nega-
tive
---
---
Pmean < –2
or
Paw < –10
PEEP/CPAP high
---
Airway pressure
>(PEEP +5 hPa
(cmH
2
O)) during more
than 10 consecutive
breaths
---
Airway pressure not
achieved
---
After 2 consecutive
breaths
---