![Dräger Atlan A300 Скачать руководство пользователя страница 76](http://html.mh-extra.com/html/drager/atlan-a300/atlan-a300_instructions-for-use-manual_2526728076.webp)
76
Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Assembly and preparation
4. Remove the absorber insert (3) from the absorber container. The sealing ring (4)
remains on the absorber insert.
Further reprocessing of the individual components is performed in accordance with
the chapter "Machine reprocessing" on page 242.
4.9.2.2
Filling and mounting
1. Push the absorber insert (5) into the absorber container (6).
2.
CAUTION
Risk of chemical burns
Soda lime is caustic and is a strong irritant for eyes, skin, and airway.
►
Handle the soda lime carefully and do not spill it.
Fill the CO
2
absorber with fresh soda lime to the upper mark.
Recommendation: Use Drägersorb 800 Plus or Drägersorb Free.
3.
WARNING
Risk of hypoventilation and incorrect gas measurement
Reuse of the disposable dust filter can increase the filter resistance and impair
the ventilation function of the device.
►
If soda lime from third-party manufacturers is used, e.g., granular soda lime,
use a disposable dust filter and replace it with every change of the soda lime.
If soda lime from third-party manufacturers is used, insert a new disposable dust
filter (7). Only use dust filters from the list of accessories.
39487
3
4
41
120
5
6