Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
275
Technical data
For internal breathing systems (restricted
range)
7 L/min
Connection for sample gas disposal when
using external patient-gas measurement
Outer hose diameter
3 to 6 mm
Maximum inlet flow
500 mL/min
16.12 Passive
AGS
Information
This system is designed for an anesthetic gas
scavenging system which works without suction
flow. Do not connect it to an active anesthetic
gas scavenging system.
To limit contamination of the ambient air in com-
pliance with ISO 80601-2-13, follow the specifi-
cations for the scavenging hose.
Scavenging hose
Maximum length
8 m (26 ft)
Minimum diameter
19 mm (0.75 in)
Connection for sample gas disposal when
using external patient-gas measurement
Outer hose diameter
3 to 6 mm
Maximum inlet flow
500 mL/min
16.13
Measuring systems and displays
Information
Alarms and internal calculations are based on
values that are more precise than the displayed
values. Consequently, there may be minor devi-
ations between the current alarm status and the
displayed measured values.
Airway pressure
Airway pressure
Paw
Plateau pressure
Pplat
Positive end-expiratory pressure
PEEP
Peak inspiratory pressure
PIP
Mean airway pressure
Pmean
Range
–20 to +99 hPa (cmH
2
O)
Accuracy
±4 % of the measured value or ±2 hPa
(cmH
2
O) (the larger value applies)
16.11 Active
AGS
(continued)