Service
14 Service
14.1 Safety
instructions
Intervals and implementation
Wear and material fatigue of the components may lead to device failure and
malfunctions. If services activities are not performed regularly or properly,
malfunctions may occur, which can result in personal injury and property damage.
►
Perform service at the specified intervals.
►
Service activities must be performed by those user groups that are assigned to
the particular measure.
►
Only perform service activities when there is no patient connected to the device.
►
Before performing maintenance, disconnect all electrical connections from the
power supply and all gas connections from the gas supply.
►
Perform a system test after service activity.
Housing
Under the housing, there are live electrical components, which may cause an
electric shock.
►
The housing may only be opened by those user groups that are assigned to that
particular measure.
Risk of infection
If the product has not been reprocessed properly, it may be contaminated with
pathogens. As a result, persons could be put at risk.
Perform the following measure before servicing, before returning the device, and
before disposal:
►
Reprocess the product in accordance with the chapter "Reprocessing".
14.2
Definition of service terminology
A service contract with Dräger is recommended.
14.3 Inspection
Concept
Definition
Service
All measures (inspection, maintenance, repair) intended to
maintain or restore the functional integrity of a product
Inspection
Measures intended to determine and assess the current state
of a product
Maintenance
Regular specified measures intended to maintain the functional
integrity of a product
Repair
Measures intended to restore the functional integrity of a prod-
uct after a failure
Measure
Interval
User group
Inspection and safety check
Every 12 months Service personnel