Tests
d. Connect the hose (6) to the coaxial test adapter (3).
2. Touch the
Tests...
button.
3. Touch the
System test
button.
The checklist opens and the breathing system pressure is displayed in the lower
left-hand corner.
4. Press the
O
2
+
key and keep it pressed until the displayed breathing system
pressure no longer rises.
5. Release the
O
2
+
key.
6. Wait until the pressure stabilizes between 15 and 30 hPa. If the pressure drops
further, the hose is leaking.
Checking the outer hose:
1. Connect the breathing circuit as follows:
a. Connect the hose (1) to the inspiratory port (2).
b. Remove the coaxial test adapter (3) from the expiratory port (4).
c. Connect the elbow (5) to the hose (6).
d. Plug the hose (6) with the elbow connected on to the circuit plug (7).
e. Connect the hose (8) to the expiratory port (4).
2. Touch the
Tests...
button.
3. Touch the
System test
button.
The checklist opens and the breathing system pressure is displayed in the lower
left-hand corner.
4. Press the
O
2
+
key and keep it pressed until the displayed breathing system
pressure no longer rises.
5. Release the
O
2
+
key.
6. Wait until the pressure stabilizes between 15 and 30 hPa. If the pressure drops
further, the hose is leaking.
7. Exit the system test again. To do this, touch the
Cancel
button.