Configuration
11.3.6.5
Vertical tab "Sound volume"
11.3.6.6
Vertical tab "Interfaces"
Headline/
Parameter
Setting range
Description
Alarm volume
10 to 100
40
Sets the alarm volume.
Minimum alarm
volume
10 to 100
10
Sets the minimum volume with which
an alarm tone will be signaled.
Headline/
Parameter
Setting range
Description
LAN
Configuration of the network
DHCP
On;
Off
Specifies settings for the network.
When using DHCP, consult with IT per-
sonnel to ensure that the device is
always assigned the same IP address
by the DHCP server.
Restart the device after each change to
the network settings. The network set-
tings are not affected by a reset to fac-
tory defaults.
Accept the changes to IP address, Sub-
net mask, or Default gateway with the
Apply button. The changes are only
active after the device has been
restarted.
IP address
XXX . XXX . XXX . XXX
Subnet mask
XXX . XXX . XXX . XXX
Default gateway
XXX . XXX . XXX . XXX
MAC address
Displays the MAC address.
COM 1
Configures the COM port.
A baud rate of 19200 or 38400 is
required for transmission of high-speed
data, e.g., for waveforms.
Protocol
MEDIBUS.X
; None
Baud rate 1200; 2400; 4800; 9600;
19200
; 38400
COM 2
Protocol
MEDIBUS.X;
None
Baud rate 1200; 2400; 4800; 9600;
19200
; 38400
USB
Activates or deactivates the USB port.
USB interface
On
; Off