Reprocessing
Absorber container
►
Place the absorber container with the opening facing down on an injector
nozzle.
CLIC adapter (option)
►
Lay the CLIC adapter in the basket with the cleaning plate facing upwards.
13.10.3 Machine
cleaning
with thermal disinfection
Use a washer-disinfector that meets the requirements of the standard ISO 15883.
Dräger recommends the use of a load carrier for anesthesia accessories and
ventilation accessories.
Follow the list of validated disinfectants (see "Validated reprocessing procedures",
page 225).
Prerequisite: The components have been positioned appropriately in the load
carrier.
1. Use a suitable cleaning agent.
2. Select a suitable cycle.
Use a cycle with the following phases:
– Pre-wash
– Cleaning using a suitable cleaning agent in accordance with the
manufacturer’s instructions for use
– Final rinsing with demineralized water
38595
38594