Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Technical data
The transferred data include the following information:
– Settings
– Measured values which comply with the specified accuracy
– Waveforms
– Text messages
– Alarm status
If the specified accuracy for measured values cannot be met, these measured
values will be displayed on the screen in gray. These measured values will not be
transmitted over MEDIBUS.X.
Take note of the documentation for the following communication protocols before
transferring the data:
– MEDIBUS.X, Rules and Standards for Implementation (9052607)
– MEDIBUS.X, Profile Definition for Data Communication V1.n (9052608)
The documents are only available in English.
The following personal data are transmitted unencrypted over the interface:
– Therapy data with details of age, weight, and height of the patient
With the aim of improving the clinical process, the data can be used to set up a
distributed alarm system with unconfirmed alarm transmission conforming to
IEC 60601-1-8. However, the data must not be used as a substitute for Atlan as the
primary alarm source.
16.25.2.1 Required
characteristics
The RS-232 port is a point-to-point connection. A connected device must prevent
access by unauthorized users to the data that are sent over the RS-232 port and
must itself be protected from infections by malware and computer viruses.
16.25.3 USB
port
The USB port supports the transfer of data to external storage media. Existing data
on external storage media may be deleted during this operation.
The following personal data are transmitted unencrypted over the interface:
– User logbook with details of age, weight, and height of the patient
– Screenshots potentially showing information about the age, weight, and height
of the patient
16.25.3.1 Required
characteristics
A connected device must conform to the mass storage medium USB device class
(e.g., connecting devices to charge the battery is not intended.). Dräger
recommends the use of FIPS 140-2 compatible storage media with hardware
encryption.
16.25.3.2 Hazardous
situations
Connecting active devices to the Atlan USB port can cause Atlan to restart.