Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Troubleshooting
!
100
Infinity ID CO
2
absorber
not compatible
An incompatible accessory is
connected.
Check accessory.
Use "ALARM RESET" to
acknowledge alarm.
!
100
Infinity ID water trap not
compatible
An incompatible accessory is
connected.
Check accessory.
Use "ALARM RESET" to
acknowledge alarm.
!!! 255
Inspiratory desflurane high Inspiratory anesthetic gas
concentration has exceeded
the upper alarm limit.
Check vaporizer and fresh-
gas settings.
!! 255
Inspiratory desflurane high Inspiratory anesthetic gas
concentration has exceeded
the upper alarm limit.
Check vaporizer and fresh-
gas settings.
!
255
Inspiratory desflurane low The inspiratory anesthetic
gas concentration is below
the lower alarm limit.
Check vaporizer and fresh-
gas settings.
Refill the vaporizer.
Check the breathing system
for high leakage.
The soda lime has dried out. Replace the soda lime.
!!! 255
Inspiratory enflurane high
Inspiratory anesthetic gas
concentration has exceeded
the upper alarm limit.
Check vaporizer and fresh-
gas settings.
!! 255
Inspiratory enflurane high
Inspiratory anesthetic gas
concentration has exceeded
the upper alarm limit.
Check vaporizer and fresh-
gas settings.
!
255
Inspiratory enflurane low
The inspiratory anesthetic
gas concentration is below
the lower alarm limit.
Check vaporizer and fresh-
gas settings.
Refill the vaporizer.
Check the breathing system
for high leakage.
The soda lime has dried out. Replace the soda lime.
!
100
Inspiratory flow sensor
expired
Accessory has been used
too long.
Replace the accessory if nec-
essary.
Use "ALARM RESET" to
acknowledge alarm.
!
255
Inspiratory flow sensor
failure
The sensor is not calibrated.
The breathing system has
been replaced or discon-
nected since last calibration.
Perform the leakage test.
Failure of the flow sensor.
Replace the flow sensor and
perform a leakage test.
!!! 255
Inspiratory halothane high Inspiratory anesthetic gas
concentration has exceeded
the upper alarm limit.
Check vaporizer and fresh-
gas settings.
!! 255
Inspiratory halothane high Inspiratory anesthetic gas
concentration has exceeded
the upper alarm limit.
Check vaporizer and fresh-
gas settings.
Priority
Alarm
Cause
Remedy