Configuration
11.3.4.5
Vertical tab "General"
11.3.5
System setup > Licenses/Options
11.3.5.1
Vertical tab "Licenses/Options"
Headline/
Parameter
Setting range
Description
Pinsp changes
with PEEP
On
; Off
Determines whether changing PEEP
automatically changes Pinsp also, so
that the difference between PEEP and
Pinsp remains constant.
Ti changes with
RR (I:E ratio is
locked)
On
; Off
Determines whether the change in
respiratory rate automatically effects a
change in Ti, so that the I:E.ratio
remains constant.
Breathing system
warmer
On
; Off
Switches the breathing system warmer
on or off.
Auto Wake-up
(opens start dia-
log upon respira-
tory activity)
On
; Off
Determines whether the Start dialog
opens automatically when ventilation
activity is detected (e.g., as a result of
repeated squeezing of the breathing
bag).
Anesthetic gas
compensation
Auto;
Off
;
Des; Sev
Corrects the accuracy of the flow mea-
surement on devices with inspiratory
O
2
measurement.
Select the value according to the typical
setting on the vaporizer control dial.
(The vaporizer automatically makes
any correction necessary for altitude.)
Des
0 to 18
6
Sev
0 to 6.5
2.1
Headline/
Parameter
Setting range
Description
Licenses for software options
– Overview of available and active
software options.
– Activating software options.
Observe the information in the following
section: "Activating software
options", page 182