Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
59
Assembly and preparation
To use the anesthetic gas compensation - see page 98 - connect an anesthetic gas
monitor to the COM 2 port.
Only use the cables from the list of accessories.
For further information on configuring the particular interface, see page 179.
4.4 Intrahospital
transport
Transport includes any movement of the device to other rooms or functional areas.
Alignment or positioning of the device within the operating room is not counted as
transport.
4.4.1
Increasing the tipping stability during transport
1. Carefully fold the arms with any mounted accessories against the device, (e.g.,
patient monitor, data management system, syringe pumps).
2. Use a strap to prevent the arms from swinging out in an uncontrolled manner.
3. Remove all loose objects from the attached arms and the shelves.
4. Remove all objects weighing more than 8 kg from the rails.
5. Remove the vaporizers.
6. Remove all objects from the writing tray and the work surface.
7. Slide the writing tray completely into the device.
8. Position the breathing bag arm (if present) close to the device.
9. Push in and lock the drawers (if present).
4.4.2 Parking
the medical device
When parking, always engage the brakes (central brake for front castors and castor
brake at rear), especially on inclined surfaces.
4.4.3
Visual inspection after transport
1. Check the device for damage, particularly the hoses and cables.
2. Do not operate a damaged device. The damage must be repaired by service
personnel before the device is used.
4.5 Gas
supply
4.5.1
Connecting to the central gas supply system
1. Screw the compressed gas hoses for the central gas supply system to the gas
inlets by hand, see page 26.
2.
WARNING
Danger to the patient and user
If the strain relief for the compressed gas hoses is not used, the device may be
damaged.
►
Use the strain relief for the compressed gas hoses.
Insert the compressed gas hoses into the strain relief. Tighten the strain relief,
see page 23.