![Dräger Atlan A300 Скачать руководство пользователя страница 106](http://html.mh-extra.com/html/drager/atlan-a300/atlan-a300_instructions-for-use-manual_2526728106.webp)
106
Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Operation
7.5.5 Volumeter
The volumeter is used for observing and assessing the ventilation.
The bar graph indicates the inspiratory and expiratory tidal volume.
At the end of the inspiration, the delivered tidal volume is displayed as a bar (1).
At the end of the expiration, the difference between inspiratory and expiratory tidal
volumes (6) is displayed.
The expiratory tidal volume is displayed next to the bar graph (3).
Using the volumeter (minute volume measurement)
1. Touch the
Start
button (2) to start.
The bar graph displays the individual measured spontaneous breaths in
segments (5). The summed volume (4) is displayed next to the bar graph.
2. Touch the
Stop
button (2) to stop.
3. To reset the volumeter and time display to zero, touch the
Reset
button (2).
The volumeter stops automatically after 60 seconds. The measured values are
displayed for 4 minutes and then deleted.
7.5.6 Low-flow
wizard
Prerequisite: The device has the "Advanced gas monitoring" option.
The low-flow wizard displays bar graphs for the required fresh-gas flow and the
current total flow. Both bar graphs are to the same scale.
An evaluation of the total flow is displayed below the bar graph:
355
53
2
3
6
4
5
1
Volumeter
VT
Volume
35554
Evaluation
Color
Meaning
Too high
Yellow
The fresh-gas flow is possibly too high. If
the fresh-gas flow can be reduced, both
fresh gas and anesthetic agent will be
saved.
Efficient
Green
No action is necessary.
Required FG flow
Total flow
Efficient