Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Troubleshooting
!! 255
Apnea
No breathing or ventilation.
Start manual ventilation!
Check ventilation settings.
Check spontaneous breath-
ing ability of the patient.
!!! 220
Apnea (no CO
2
)
No breathing or ventilation.
Start manual ventilation!
Check ventilation settings.
Check spontaneous breath-
ing ability of the patient.
Sample line is not con-
nected.
Connect sample line to
breathing circuit.
!! 255
Apnea (no CO
2
)
No breathing or ventilation.
Start manual ventilation!
Check ventilation settings.
Check spontaneous breath-
ing ability of the patient.
Sample line is not con-
nected.
Connect sample line to
breathing circuit.
!!! 220
Apnea (no flow)
No breathing or ventilation.
Start manual ventilation!
Check ventilation settings.
Check spontaneous breath-
ing ability of the patient.
Fresh-gas flow is insuffi-
cient, breathing bag is
blocked or positioned incor-
rectly.
Check fresh-gas settings and
position of breathing bag.
The breathing hoses are
blocked or leaking.
Check breathing circuit and
tube.
!! 255
Apnea (no flow)
No breathing or ventilation.
Start manual ventilation!
Check ventilation settings.
Check spontaneous breath-
ing ability of the patient.
Fresh-gas flow is insuffi-
cient, breathing bag is
blocked or positioned incor-
rectly.
Check fresh-gas settings and
position of breathing bag.
The breathing hoses are
blocked or leaking.
Check breathing circuit and
tube.
!!! 220
Apnea (no pressure)
No breathing or ventilation.
Start manual ventilation!
Check ventilation settings.
Fresh-gas flow is insuffi-
cient, breathing bag is
blocked or positioned incor-
rectly.
Check fresh-gas settings and
position of breathing bag.
The breathing hoses are
blocked or leaking.
Check breathing circuit and
tube.
Priority
Alarm
Cause
Remedy