Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
291
Technical data
16.21.2
Response times in event of concentration changes
Typical response times (T
0..90
) for an oxygen concentration change from 21 Vol% to
100 Vol% at the following fresh-gas flows:
16.22 EMC
declaration
16.22.1 General
information
This device was tested for electromagnetic compatibility using accessories from the
list of accessories. Other accessories may only be used if they do not compromise
the electromagnetic compatibility. The use of non-compliant accessories may result
in increased electromagnetic emissions or decreased electromagnetic immunity of
the device.
This device may be used in the direct vicinity of other devices only if Dräger has
approved this device arrangement. If no approval has been given by Dräger, it must
be ensured that this device functions correctly in the desired arrangement before
use. The instructions for use for the other devices must be followed.
With
breathing circuit for
adults
MP00349, inspiratory filter
MP01730
60
–5.6
5.5
30
–2.0
3.4
With
breathing circuit for
pediatric
patients
MP01340 and MP01343,
inspiratory filter MP01815
15
–1.6
3.5
With
breathing circuit for
neonates
MP00333, inspiratory filter
MP01815
2.5
0.6
1.8
Breathing system, dry, with filled
reusable CO
2
absorber and soda
lime dust filter MX50115
Peak flow in
use [L/min]
Resistance [hPa (cmH
2
O)]
Man / Spon
Inspiratory
Expiratory
2 L/min
4 L/min
8 L/min
O
2
flush
Test lung for adults
(MP02400), breathing circuit
(MP00300), breathing bag 2 L
(MP00222)
VT
=500 mL,
RR
=10 /min,
I:E
=1:2
712 s
174 s
32 s
9 s
Test lung for neonates
(8410079), breathing circuit
(MP00333), breathing bag 1 L
(MP00383)
VT
=30 mL,
RR
=30 /min,
I:E
=1:1
91 s
64 s
46 s
7 s