Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Troubleshooting
4.
CAUTION
Risk of contaminating the ambient air
A filter cover that is not properly closed can cause contamination of the ambient
air with anesthetic gases.
►
Take care that the cover is properly closed.
Fold up the cover.
12.10.2 Replacing
the
anesthetic gas receiving system (AGS)
This chapter describes the replacement using the example of an active AGS which
is connected to the device with the transfer hose and to the anesthetic gas
scavenging system with the scavenging hose.
Disassembly
1. Disconnect the transfer hose (1).
2. Disconnect the scavenging hose (2).
3. Unfasten the 2 screws (3).
4. Withdraw the AGS to the side.
5. Dispose of the AGS.
The AGS must not be reprocessed.
Assembly
1. Slide in the new AGS.
2. Fasten the AGS with 2 screws.
3. Connect the scavenging hose.
4. Connect the transfer hose.