Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Principles of operation
17.2.6 Volume-controlled
ventilation
17.2.6.1 Compliance compensation
The applied VT is corrected by the determined breathing hose compliance, i.e., an
additional volume is delivered in order to ensure the application of the volume to the
patient. On the product variant with the integrated patient-gas measurement
module, the applied VT is corrected by the amount of suction flow as soon as CO
2
respiratory phases are detected.
17.2.6.2 VC
17.2.6.3
VC - CMV
– Volume-controlled
– Pressure-limited
– Time-controlled
– Machine-triggered
– Constant inspiratory flow
In this volume-controlled ventilation mode, the patient receives the set tidal volume
VT with each mandatory breath.