Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
95
Operation
7.2.2
Checking the patient data
►
WARNING
Risk due to incorrect settings
Different standard alarm limits or therapy settings might be configured for
medical devices within the same area. The user must observe the following:
►
Make sure that the set values and alarm limits are selected to suit the patient.
►
Make sure that the alarm system is neither rendered useless by setting
extreme values for the alarm limits nor deactivated by switching off the
alarms.
►
Check the therapy settings, the start settings for alarms, and the alarm
settings during a change of ventilation mode.
►
Only turn off alarms if the safety of the patient will not be compromised as a
result.
CAUTION
Risk due to incorrect setting for patient age
Incorrectly setting the patient age can lead to incorrect xMAC values and thus to
an incorrect anesthetic agent delivery.
►
Always set the patient age correctly.
Adjust the patient data (1).
The device will suggest appropriate therapy settings for these data, e.g., for tidal
volume, respiratory rate, and alarm limits. For more information, see page 311.
7.2.3
Setting and starting the therapy
1. Select the ventilation mode (1).
The following ventilation modes are available:
–
Man / Spon
–
VC
–
PC
–
VC - AF
–
PSV
35298
1
35300
1
2
5
4
3
Select mode