Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
191
Troubleshooting
12.1.2 Systematic
localization of leakage
To find causes for leakage, isolate individual components from the leakage test.
1. Perform the leakage test, see page 127.
2. Contact service personnel if the leakage cannot be localized.
12.1.3
Automatic leakage compensation
The device is able to compensate for small leakage.
– The pressure is held constant at the PEEP level.
– With pressure-controlled ventilation, the pressure is regulated according to the
set pressure. Small leakage is compensated by the piston drive.
12.2
Power supply failure
12.2.1
Mains power supply failure
If mains power supply fails, the device automatically switches to the internal battery.
A fully charged battery will maintain operation for at least 45 minutes. For further
information see: "Technical data", page 267.
Remaining battery charge is displayed on the status display.
The breathing system warmer is deactivated during battery operation. The peak
inspiratory flow may be limited to 75 L/minute.
12.2.2
Mains power supply failure and battery discharged
If mains power supply fails and the battery is discharged, a signal tone is emitted.
Manual ventilation and spontaneous breathing are still available. O
2
and anesthetic
agent can still be delivered using the emergency O
2
delivery (with electronically
controlled gas mixer) or the flow control valves (with mechanically controlled gas
mixer) and the vaporizers.
The following components and functions are not available:
– Ventilator
– Electronically controlled gas mixer
– Device monitoring and patient monitoring
Component
Measure
Sample line
Remove the sample line and seal the Luer-Lock connector
on the Y-piece.
Breathing hoses
Disconnect the breathing hoses. Connect the inspiratory
port and expiratory port with a hose that is known to be with-
out leakage. Connect the breathing bag directly to the
breathing system.
Vaporizers
Remove the vaporizers.
O
2
sensor
Remove the O
2
sensor. Seal the sensor port with the sealing
cap.