Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
203
Troubleshooting
!!! 110
Air supply low
Central supply pressure and
cylinder pressure are low.
Check central Air supply or
use cylinder.
Use "ALARM RESET" to
downgrade alarm priority.
!!! 255
Airway press. continu-
ously high
Airway pressure has been
continuously high.
Check spontaneous breath-
ing ability of the patient.
Check ventilation settings.
Check breathing hoses,
breathing system, and anes-
thetic gas scavenging sys-
tem.
In Man/Spon mode, check
the APL valve setting.
Check lower alarm limit for
airway pressure.
!!! 255
Airway pressure high
The upper alarm limit for the
airway pressure has been
exceeded. The applied inspi-
ratory pressure is higher than
the set value.
Check patient condition.
Check ventilation settings.
Check alarm limit.
Breathing hoses are blocked
or the tube is kinked.
Check breathing circuit and
tube.
!!! 255
Airway pressure negative
Fresh-gas flow is insuffi-
cient, breathing bag is
blocked or positioned incor-
rectly.
Check fresh-gas settings and
position of breathing bag.
Suction maneuver during
ventilation.
Check the bronchial suction
system.
Failure of the anesthetic gas
scavenging system.
Check anesthetic gas scav-
enging system.
!! 90
Airway pressure not
achieved
Fresh-gas flow is insuffi-
cient, breathing bag is
blocked or positioned incor-
rectly.
Check fresh-gas settings and
position of breathing bag.
Leakage or disconnection.
Check the breathing circuit
for tight connections and
leakages.
!
100
Alarm silence key stuck
Key is stuck or was pressed
for more than 10 seconds.
The therapy will continue with
current settings.
If the problem persists, call
Dräger.
!!! 220
Apnea
No breathing or ventilation.
Start manual ventilation!
Check ventilation settings.
Check spontaneous breath-
ing ability of the patient.
Priority
Alarm
Cause
Remedy