Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Configuration
11.3.3.4
Vertical tab "Config. 1"
Paw
+[mbar]; [hPa];
[cmH
2
O]
Off; 5 to 20
5
The following parameters are taken into
account when determining the Paw
value: PIP, Pplat, Pinsp, PEEP, and
∆
Psupp.
In the Man / Spon and Pause modes,
the new alarm limit is at least 25 hPa
(cmH
2
O).
MV ± [%]
Off; 20 to 80
40
In modes with low or no breathing sup-
port (Man / Spon, CPAP / PSV, and
Pause), a further 20 percentage points
are added to the configured value.
Headline/
Parameter
Setting range
Description
General alarm behavior
Open "Quick
setup" when
alarm occurs
On
; Off
Automatically opens the Quick setup
dialog in the event of an alarm.
"Second agent
detected" alarm
On
; Off
Issues an alarm when an anesthetic
gas mixture is detected.
"xMAC low" alarm
On
; Off
Activates the xMAC low alarm.
"FiO
2
too high for
neonates" thresh-
old value [%]
50
; 25 to 90
Sets the FiO
2
value that is considered
critical for neonates. If this value is
exceeded for a certain time, the FiO
2
too high for neonates alarm will be trig-
gered.
"FiO
2
too high for
neonates" alarm
after [h:mm]
Off; 0:10 to 9:50
0:15
Sets the time after which the FiO
2
too
high for neonates alarm will be trig-
gered.
Priority of "Apnea
ventilation" alarm
Medium
; Low
Specifies the alarm priority when the
set minimum respiratory rate is not
reached in PSV ventilation mode.
Alarm behavior in "Pause" mode
Priority of "Pause
time expired"
alarm
High
; Medium; Low
Specifies the alarm priority for the alarm
that is issued when the time set in the
Pause mode has expired.
Default value for
"Timer" [mm:ss]
0:30 to 2:00
2.00
0:30 to 2:00
1:00
0:30 to 2:00
0:30
Specifies the default duration for
Pause.
Headline/
Parameter
Setting range
Description