Troubleshooting
!!! 110
No Air delivery
Air is not available. Gas
mixer is using 100 % O
2
instead.
Check central Air supply or
use cylinder.
Use "ALARM RESET" to
downgrade alarm priority.
!
255
No CO
2
detected
Ventilation was started, but
no exhaled CO
2
was
detected for more than 60
seconds.
Check patient condition.
Check sample line, water
trap, and patient-side filter.
Use "ALARM RESET" to
acknowledge alarm.
!!! 200
No fresh-gas flow
No fresh-gas flow is set.
Open the flow control valves.
Use "ALARM RESET" to
downgrade alarm priority.
!!! 110
No N
2
O delivery
N
2
O is not available. Gas
mixer is using O
2
instead.
Check central N
2
O supply or
use cylinder.
Use "ALARM RESET" to
downgrade alarm priority.
!!! 210
No O
2
delivery
O
2
is not available. Gas
mixer is using Air instead.
Check central O
2
supply or
use cylinder.
Use "ALARM RESET" to
downgrade alarm priority.
!
255
O
2
concentration implausi-
bly high
The O
2
sensor calibration is
invalid.
Use substitute O
2
monitoring.
Perform the system test.
!
220
O
2
cylinder almost empty
The cylinder is almost empty. Replace the cylinder. Use the
central supply.
!!! 210
O
2
cylinder empty
The cylinder is empty or
closed.
Replace the cylinder. Use the
central supply.
Use "ALARM RESET" to
downgrade alarm priority.
!
255
O
2
cylinder empty
The cylinder is empty or
closed.
Replace the cylinder. Use the
central supply.
!
190
O
2
cylinder sensor?
Cylinder pressure sensor is
not connected.
Check if cylinder pressure
sensor is connected.
Use "ALARM RESET" to
acknowledge alarm.
!! 150
O
2
FG flow measurement
failed
The measurement system for
the O
2
fresh-gas flow has
failed.
Only use O
2
as fresh gas.
Use the total flow tube to
check the fresh-gas flow.
Adjust the fresh-gas flow so
that it equals or exceeds the
minute volume.
Priority
Alarm
Cause
Remedy