Technical data
Ambient pressure
500 to 1060 hPa
(7.3 to 15.3 psi)
Relative humidity
10 to 95 %, non-condensing
CO
2
concentration
Not relevant
Information
The permissible ambient conditions depend on
the accessories used. Follow the corresponding
instructions for use.
16.4 Fresh-gas
delivery
Gas mixer
The device is equipped with either an electroni-
cally controlled or a mechanically controlled gas
mixer.
Standardization
All data are standardized to STPD conditions.
O
2
flush
25 to 75 L/min at 2.7 to 6.9 kPa x 100 (39 to
100 psi; 0.27 to 0.69 MPa) supply pressure
O
2
flow with integrated flowmeter
Range
Off; 2 to at least 10 L/min at 2.7 kPa x 100
(39 psi or 0.27 MPa) supply pressure
Accuracy
±10 % of the set value for flows >2.0 L/min
Resolution of displayed value
1 L/min (up to 10 L/min)
5 L/min (starting at 10 L/min)
O
2
flow with external flowmeter
Range
Off; 1 to at least 10 L/min at 2.7 kPa x 100
(39 psi or 0.27 MPa) supply pressure
Accuracy
±10 % of the set value
Resolution of displayed value
1 L/min (up to 10 L/min)
5 L/min (starting at 10 L/min)
16.5
Fresh-gas delivery with electronically controlled gas
mixer
Standardization
All data are standardized to STPD conditions.
O
2
concentration FG O
2
Setting range
21 to 100 Vol% (carrier gas: Air)
25 to 100 Vol% (carrier gas: N
2
O)
Accuracy
±5 % or ±2 Vol% (the larger value applies)
Fresh-gas flow FG flow
Setting range
Off; 0.2 to 15 L/min
16.3 Ambient