Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
143
Tests
8.4.4 Valves
8.4.4.1
Checking the O
2
flush
1. Press the
O
2
+
key and keep it pressed until the pressure stabilizes.
The pressure does not exceed 45 hPa (cmH
2
O).
2. Release the
O
2
+
key again and wait until the pressure stabilizes.
The pressure does not fall below 15 hPa (cmH
2
O).
3. Lift the APL valve.
The pressure is released.
351
18
351
18
35120
20
40