Operation
This button is only visible in the following ventilation modes:
–
Manual / Spontaneous
–
External fresh-gas outlet
–
Pause
The alarm system is immediately activated when the CO
2
monitoring is activated.
Deactivation is indicated in the header bar and in the parameter field by the
symbol.
7.9.3
Automatic xMAC monitoring
Prerequisite: The device has the "Integrated patient-gas measurement module"
option.
The xMAC monitoring is automatically activated as soon as the following conditions
are met:
– Anesthetic gas is administered.
– The inspiratory xMAC value is greater than the expiratory xMAC value.
– The expiratory xMAC value reaches approximately 0.3.
If the xMAC value rises, the xMAC monitoring (1) is activated and the lower alarm
limit for the xMAC level is automatically adapted to the anesthetic gas
concentration. The lower alarm limit (2) can reach a maximum value of 1.0.
The lower alarm limit can be recalculated by touching the
Autoset
button (3). This
allows the alarm limit for the
xMAC low
alarm to be adjusted in special anesthesia
situations and to exceed the value of 1.0 if necessary.
If the expiratory xMAC value falls below the alarm limit (4 or 7), the device issues
the
xMAC low
alarm with low priority. If the alarm is not acknowledged with the
ALARM RESET
button, the priority is raised to medium priority after 60 seconds.