4 Preparation for Operation (Startup)
4.1 Overview of Actions
Important:
The pump is filled with 2-propanol when being shipped from the
factory. During initial operation of the module, make sure that the
solvents used are miscible with 2-propanol. Otherwise, follow the
appropriate intermediate steps.
Important:
La pompe est stockée sous 2-propanol lorsqu'elle est expédiée depuis
l'usine. Lors du démarrage initial de la pompe, assurez-vous que les
solvants utilisés sont miscibles. Dans le cas contraire, utilisez un
solvant intermédiaire approprié. Même d'infimes particules peuvent
endommager le système.
After you have unpacked, positioned and connected the pump as described in sections 3.1
through 3.4 (
page 27 and following pages), prepare the pump for operation. Follow this
sequence of steps:
1.
If the UltiMate 3000 system includes an SRD-3x00 Solvent Rack
Connect the 15-pin D-Sub port on the rear of the Solvent Rack with the Solvent Rack port
on the rear of the pump (
2.
If you want to connect an ESA Corona or Coulochem III detector to the pump
Refer to section 3.6 for details (
Before
you attach the column and cell, flush the system with several volumes of the
mobile phase to be used.
3.
Connect the pump to the other modules of your UltiMate 3000 system, as required by
your application.
When you connect capillaries to the pump, observe the general precautions on page 47.
4.
Connect the solvent reservoirs, and then connect the solvent lines to the degassing
channels (
5.
Set up the rear seal wash system (
page 50) and flush the system with the seal-
washing liquid (
page 51).
6.
Connect drain tubing (
This is a
must
as also the seal washing liquid is directed to waste through the drain
system. If drain tubing is not connected, modules that are located below the pump in the
UltiMate 3000 system stack, may suffer severe damage from the liquid leaving the drain
port.
Содержание UltiMate 3000 Series
Страница 2: ...UltiMate 3000 Series SD RS and BM Pumps Operating Instructions ...
Страница 4: ...UltiMate 3000 Series SD RS and BM Pumps Page II Operating Instructions ...
Страница 5: ...UltiMate 3000 Series SD RS and BM Pumps Operating Instructions Page III ...
Страница 6: ...UltiMate 3000 Series SD RS and BM Pumps Page IV Operating Instructions ...
Страница 7: ...UltiMate 3000 Series SD RS and BM Pumps Operating Instructions Page V ...
Страница 8: ...UltiMate 3000 Series SD RS and BM Pumps Page VI Operating Instructions ...
Страница 14: ...UltiMate 3000 Series SD RS and BM Pumps Page vi Operating Instructions ...
Страница 24: ...UltiMate 3000 Series SD RS and BM Pumps Page 10 Operating Instructions ...
Страница 58: ...UltiMate 3000 Series SD RS and BM Pumps Page 44 Operating Instructions ...
Страница 74: ...UltiMate 3000 Series SD RS and BM Pumps Page 60 Operating Instructions ...
Страница 116: ...UltiMate 3000 Series SD RS and BM Pumps Page 102 Operating Instructions ...
Страница 136: ...UltiMate 3000 Series SD RS and BM Pumps Page 122 Operating Instructions ...
Страница 200: ...UltiMate 3000 Series SD RS and BM Pumps Page 186 Operating Instructions ...
Страница 204: ...UltiMate 3000 Series SD RS and BM Pumps Page 190 Operating Instructions ...
Страница 228: ...UltiMate 3000 Series SD RS and BM Pumps Page 214 Operating Instructions ...
Страница 236: ...UltiMate 3000 Series SD RS and BM Pumps Page 222 Operating Instructions ...