UltiMate 3000 Series:
SD, RS, and BM Pumps
Page
92
Operating
Instructions
After you have …
Perform the following command …
Replaced the entire pump head assembly CheckValveServiceDone
1
, PistonsChanged
1
,
SealChanged
1
, SupportRingChanged
1
Replaced a support ring SupportRingChanged
1
Serviced the pump (for example, annual maintenance)
ServiceDone
3
Performed instrument qualification
QualificationDone
3
1 Listed in the Commands dialog box under PumpModule > [Pump Device Name]_Wellness_LeftBlock
and/or PumpModule > [Pump Device Name]_Wellness_RightBlock
2 Listed in the Commands dialog box under PumpModule > [Pump Device Name]_Wellness
3
Listed in the Commands dialog box under PumpModule_Wellness
For information about the [Pump Device Name], see page 37.
These commands reset the related counters and update the information when the action
was performed.
Control Panel
On the control panel for the pump, click
Wellness
,
Qualification
, and
Service
to see the
related predictive performance commands and parameters on separate panels. On these panels,
you can enter the limits and reset the counters. In addition, wellness bars provide visual
indicators of qualification and service periods. The color-coding of the wellness bars provides
information about the status:
Color
Description
Green OK.
Yellow
The value will soon reach the specified limit and/or the related component needs servicing
or should be replaced soon.
Orange
(Only for monitoring Qualification properties.) The value has reached the specified limit.
However, a Grace Period has been specified during which the pump may still be operated.
Rot
The value has reached the specified limit or the specified grace period has expired.
Replacement of a component, servicing, or qualification of the pump is overdue. The pump
can no longer be operated; besides, it is not possible to start a batch.
In addition, a message appears in the Chromeleon Audit Trail when a limit has been
reached.
Содержание UltiMate 3000 Series
Страница 2: ...UltiMate 3000 Series SD RS and BM Pumps Operating Instructions ...
Страница 4: ...UltiMate 3000 Series SD RS and BM Pumps Page II Operating Instructions ...
Страница 5: ...UltiMate 3000 Series SD RS and BM Pumps Operating Instructions Page III ...
Страница 6: ...UltiMate 3000 Series SD RS and BM Pumps Page IV Operating Instructions ...
Страница 7: ...UltiMate 3000 Series SD RS and BM Pumps Operating Instructions Page V ...
Страница 8: ...UltiMate 3000 Series SD RS and BM Pumps Page VI Operating Instructions ...
Страница 14: ...UltiMate 3000 Series SD RS and BM Pumps Page vi Operating Instructions ...
Страница 24: ...UltiMate 3000 Series SD RS and BM Pumps Page 10 Operating Instructions ...
Страница 58: ...UltiMate 3000 Series SD RS and BM Pumps Page 44 Operating Instructions ...
Страница 74: ...UltiMate 3000 Series SD RS and BM Pumps Page 60 Operating Instructions ...
Страница 116: ...UltiMate 3000 Series SD RS and BM Pumps Page 102 Operating Instructions ...
Страница 136: ...UltiMate 3000 Series SD RS and BM Pumps Page 122 Operating Instructions ...
Страница 200: ...UltiMate 3000 Series SD RS and BM Pumps Page 186 Operating Instructions ...
Страница 204: ...UltiMate 3000 Series SD RS and BM Pumps Page 190 Operating Instructions ...
Страница 228: ...UltiMate 3000 Series SD RS and BM Pumps Page 214 Operating Instructions ...
Страница 236: ...UltiMate 3000 Series SD RS and BM Pumps Page 222 Operating Instructions ...