UltiMate 3000 Series:
SD, RS, and BM Pumps
Operating Instructions
Page
105
6.3 Messages on the Pump Display
Each time a fault or error occurs during the operation of the pump, the
Status
LED on the
front panel door changes to red and a message appears on the pump display. In this case,
the
Prev
,
Next
, and
Clear
keys appear on the navigation bar.
To …
Select ...
Return to the previous message.
Prev
Proceed to the next message.
Next
Remove all messages from the display.
Clear
These keys are active also when the pump is connected in Chromeleon.
When the pump is operated with Chromeleon
The error is also displayed in the Chromeleon Audit Trail.
Messages on the pump display can be removed also by performing the
ClearDisplayError
command in Chromeleon.
The table on page 105 lists pump-related messages and suggests appropriate remedial
actions. The component to which the message belongs may appear in front of the message
text.
If the pump is operated with Chromeleon, consider also checking the Audit Trail message.
The Audit Trail may provide additional information.
In addition to the messages listed in the table, other messages may appear. In this case, note
the exact wording of the message and contact Dionex Service if you are unable to eliminate
the problem.
Error Message
Remedial Action
(x) counts deviation in zero
position
(where x is the exact deviation from the zero position.)
Turn the pump off and on again by pressing the power switch on the
rear of the pump. If the message appears again, contact Dionex
Service.
All flows need to be stopped
before servicing the pistons.
An attempt was made to move the piston into the position for piston
or piston seal replacement although the flow rate was not yet down to
zero.
Set the pump flow rate to 0. Wait until the system pressure is down to
zero.
An emergency stop was
requested over the digital
input line
This message appears if an ESA Corona detector is connected to the
pump and the digital input is configured to stop the pump when an
error occurs in the detector. Check the detector and take appropriate
remedial action (
Detector manual
).
Содержание UltiMate 3000 Series
Страница 2: ...UltiMate 3000 Series SD RS and BM Pumps Operating Instructions ...
Страница 4: ...UltiMate 3000 Series SD RS and BM Pumps Page II Operating Instructions ...
Страница 5: ...UltiMate 3000 Series SD RS and BM Pumps Operating Instructions Page III ...
Страница 6: ...UltiMate 3000 Series SD RS and BM Pumps Page IV Operating Instructions ...
Страница 7: ...UltiMate 3000 Series SD RS and BM Pumps Operating Instructions Page V ...
Страница 8: ...UltiMate 3000 Series SD RS and BM Pumps Page VI Operating Instructions ...
Страница 14: ...UltiMate 3000 Series SD RS and BM Pumps Page vi Operating Instructions ...
Страница 24: ...UltiMate 3000 Series SD RS and BM Pumps Page 10 Operating Instructions ...
Страница 58: ...UltiMate 3000 Series SD RS and BM Pumps Page 44 Operating Instructions ...
Страница 74: ...UltiMate 3000 Series SD RS and BM Pumps Page 60 Operating Instructions ...
Страница 116: ...UltiMate 3000 Series SD RS and BM Pumps Page 102 Operating Instructions ...
Страница 136: ...UltiMate 3000 Series SD RS and BM Pumps Page 122 Operating Instructions ...
Страница 200: ...UltiMate 3000 Series SD RS and BM Pumps Page 186 Operating Instructions ...
Страница 204: ...UltiMate 3000 Series SD RS and BM Pumps Page 190 Operating Instructions ...
Страница 228: ...UltiMate 3000 Series SD RS and BM Pumps Page 214 Operating Instructions ...
Страница 236: ...UltiMate 3000 Series SD RS and BM Pumps Page 222 Operating Instructions ...