Operating
Instructions
Do not place any heavy objects on the open front panel door. This may damage the door.
Always set a lower pressure limit for the HPLC pump. This prevents damage resulting
from leakage or from running the pump dry over a longer period. Activate liquid level
monitoring for the solvent reservoirs (
To avoid that the pressure calibration of the pump is impaired, turn on the pump only
when the pump pressure is down. To ensure that the pressure is down, open the purge
valve before turning on the pump. If you are operating an ISO-3100BM, observe the
precautions on page 90.
Never run the pump dry. Damage to the pistons or the piston seals could result.
Before you start operating the pump, check the liquid level in the wash liquid reservoir
and refill the reservoir as needed. After turning on the pump, wait until the wash liquid
has passed all pump heads.
Always use fresh rear seal wash liquid.
Dionex advises against recycling the solvents. This may impair the performance of the
seals.
When connecting the capillaries, make sure that the connectors are free from
contaminants. Even minute particles may cause damage to the system, for example, to the
column.
After operation, rinse out buffers and solutions that form peroxides.
Before switching from buffer to organic solution, rinse the pump thoroughly with
deionized water.
When switching to another solvent, ensure that the new solvent is miscible with the one
contained in the pump. Otherwise, the pump can be damaged; for example, by
flocculation.
If the pump flow is interrupted for longer periods (> 1 hour), turn off the lamps in any UV
or RF detector connected to the pump. This will prevent evaporation in the flow cell.
If you want to connect an ESA Corona or Coulochem III detector to the pump, refer to
page 43 for details.
Always use the frits recommended by Dionex to prevent particulate matters from entering
the HPLC system. Using other frits may considerably affect the system performance.
Do not use stainless steel frits with the BM pumps; this renders the biocompatibility void.
Frits are used on the drawing side (in the solvent reservoirs) and on the high-pressure side
(in the inline filter).
If the mobile phase includes volatile or flammable solvents, avoid open flames and
sparks.
Содержание UltiMate 3000 Series
Страница 2: ...UltiMate 3000 Series SD RS and BM Pumps Operating Instructions ...
Страница 4: ...UltiMate 3000 Series SD RS and BM Pumps Page II Operating Instructions ...
Страница 5: ...UltiMate 3000 Series SD RS and BM Pumps Operating Instructions Page III ...
Страница 6: ...UltiMate 3000 Series SD RS and BM Pumps Page IV Operating Instructions ...
Страница 7: ...UltiMate 3000 Series SD RS and BM Pumps Operating Instructions Page V ...
Страница 8: ...UltiMate 3000 Series SD RS and BM Pumps Page VI Operating Instructions ...
Страница 14: ...UltiMate 3000 Series SD RS and BM Pumps Page vi Operating Instructions ...
Страница 24: ...UltiMate 3000 Series SD RS and BM Pumps Page 10 Operating Instructions ...
Страница 58: ...UltiMate 3000 Series SD RS and BM Pumps Page 44 Operating Instructions ...
Страница 74: ...UltiMate 3000 Series SD RS and BM Pumps Page 60 Operating Instructions ...
Страница 116: ...UltiMate 3000 Series SD RS and BM Pumps Page 102 Operating Instructions ...
Страница 136: ...UltiMate 3000 Series SD RS and BM Pumps Page 122 Operating Instructions ...
Страница 200: ...UltiMate 3000 Series SD RS and BM Pumps Page 186 Operating Instructions ...
Страница 204: ...UltiMate 3000 Series SD RS and BM Pumps Page 190 Operating Instructions ...
Страница 228: ...UltiMate 3000 Series SD RS and BM Pumps Page 214 Operating Instructions ...
Страница 236: ...UltiMate 3000 Series SD RS and BM Pumps Page 222 Operating Instructions ...