UltiMate 3000 Series:
SD, RS, and BM Pumps
Page
138
Operating
Instructions
4.
Reconnect the fluid connections on the pump head. Observe the following:
When installing the capillary from the pump head to the purge unit, make sure that you
install the capillary in the direction of solvent flow, as indicated by the arrow on the
capillary. If the label is missing: Note that one side of the capillary is longer than the
other. Connect the longer side to the pump head.
When you attach the solvent line to the inlet valve, take care to avoid cross-threading.
5.
Only HPG-3x00
Reconnect the pressure transducer cable to the P-Work connector on the interior front
panel.
Verify that the cable is properly connected. If it is not, messages relating the working
pressure may appear on the pump display, for example, 'Working piston pressure out of
range.'
6.
Return the pistons into the appropriate position for normal operation. To do so,
perform the appropriate command for the pump head that you want to install:
On the
Pump/Maintenance
screen on the pump display, select
Dock
. Wait until the
screen returns to the Main menu before continuing with the next steps.
—or—
In Chromeleon, perform the
DockPistons
command from the
Commands
dialog box
(listed under
[Pump Device Name]_Wellness_RightBlock
and/or
[Pump Device
Name]_Wellness_LeftBlock
; for information about the [Pump Device Name], see
page 37). Wait until the
PistonPositionStatus
property reports
Operational
before
you continue with the next steps.
7.
Flush the pump head with seal-washing liquid (
8.
Rinse the pump thoroughly to prevent contaminants from entering the HPLC system.
Rinse the pump with at least 30 mL of HPLC-grade water or pure organic solution.
Open the purge valve (with an ISO-3100BM, observing the precautions on page 90) to
prevent the rinsing liquid from entering the HPLC system.
9.
Test the pump for leakage (
page 154). Tighten leaking connections.
10.
Update the related service information in Chromeleon. To do so, perform the
PistonsChanged
command for the pump head for which you replaced the pistons
(
page 91).
If you replaced the entire pump head assembly, perform also the
SealChanged
,
CheckValveServiceDone
, and
SupportRingChanged
commands.
Содержание UltiMate 3000 Series
Страница 2: ...UltiMate 3000 Series SD RS and BM Pumps Operating Instructions ...
Страница 4: ...UltiMate 3000 Series SD RS and BM Pumps Page II Operating Instructions ...
Страница 5: ...UltiMate 3000 Series SD RS and BM Pumps Operating Instructions Page III ...
Страница 6: ...UltiMate 3000 Series SD RS and BM Pumps Page IV Operating Instructions ...
Страница 7: ...UltiMate 3000 Series SD RS and BM Pumps Operating Instructions Page V ...
Страница 8: ...UltiMate 3000 Series SD RS and BM Pumps Page VI Operating Instructions ...
Страница 14: ...UltiMate 3000 Series SD RS and BM Pumps Page vi Operating Instructions ...
Страница 24: ...UltiMate 3000 Series SD RS and BM Pumps Page 10 Operating Instructions ...
Страница 58: ...UltiMate 3000 Series SD RS and BM Pumps Page 44 Operating Instructions ...
Страница 74: ...UltiMate 3000 Series SD RS and BM Pumps Page 60 Operating Instructions ...
Страница 116: ...UltiMate 3000 Series SD RS and BM Pumps Page 102 Operating Instructions ...
Страница 136: ...UltiMate 3000 Series SD RS and BM Pumps Page 122 Operating Instructions ...
Страница 200: ...UltiMate 3000 Series SD RS and BM Pumps Page 186 Operating Instructions ...
Страница 204: ...UltiMate 3000 Series SD RS and BM Pumps Page 190 Operating Instructions ...
Страница 228: ...UltiMate 3000 Series SD RS and BM Pumps Page 214 Operating Instructions ...
Страница 236: ...UltiMate 3000 Series SD RS and BM Pumps Page 222 Operating Instructions ...