UltiMate 3000 Series:
SD, RS, and BM Pumps
Page
106
Operating
Instructions
Error Message
Remedial Action
Can't do this when the flow
is on.
An attempt was made to perform a self test although the flow rate
was not yet down to zero.
Set the pump flow rate to 0 and repeat the command.
Can't reset pressure sensor.
Pressure is not constant.
The pump pressure is not down to zero during the self test. Wait until
the pressure is down and repeat the test. Verify that the cable for the
system pressure transducer is properly connected to the P-Sys
connector on the interior front panel.
Can't reset pressure sensor.
Pressure is out of range.
The pump pressure is not down to zero during the self test. Wait until
the pressure is down and repeat the test. Verify that the cable for the
system pressure transducer is properly connected to the P-Sys
connector on the interior front panel.
Can't start pump while alarm
is on.
An alarm has occurred, for example, a leak alarm. You can restart the
pump flow only after having acknowledged the alarm, for example,
by performing the AlarmOff command in Chromeleon.
Compression limit reached
In addition, the pump block status LED flashes green.
The compression value was 100% during each of the past three
strokes. If the compression value is lower than 100% for several
strokes afterward, the message 'Compression back to normal' will
appear on the display.
For more information about the compression values, see section 6.6
(
page 121).
Configuration error: Found
x
ID chips and
y
pump blocks.
(where x and y are numbers)
A pump block is defective. Contact Dionex Service.
Degasser malfunction
LPG-3400
The vacuum level monitoring function of the degasser in the pump
recognized insufficient vacuum. Turn the pump off and on again by
pressing the power switch on the rear of the pump. Check the
degasser vacuum in Chromeleon (Commands dialog box). After
about 1 minute, the setting should change from NotOk to OK. Inspect
the degassing module for indications of leakage it the vacuum is still
insufficient. The degassing module may be defective and should be
replaced by Dionex service.
Pumps to which an SRD-3x00 Solvent Rack is connected
The vacuum level monitoring function of the degasser in the Solvent
Rack was activated. The vacuum level is insufficient.
The degasser may not be connected properly to the pump. Verify that
the Solvent Rack is properly connected to the pump. Turn the Solvent
Rack off and on again by pressing the standby button on the front
panel. The degassing module may be defective and should be
replaced by Dionex service.
I2C device
xx
does not work
(where xx is a number indicating the device. The number is important
for servicing.)
Turn the pump off and on again by pressing the power switch on the
rear of the pump.
Содержание UltiMate 3000 Series
Страница 2: ...UltiMate 3000 Series SD RS and BM Pumps Operating Instructions ...
Страница 4: ...UltiMate 3000 Series SD RS and BM Pumps Page II Operating Instructions ...
Страница 5: ...UltiMate 3000 Series SD RS and BM Pumps Operating Instructions Page III ...
Страница 6: ...UltiMate 3000 Series SD RS and BM Pumps Page IV Operating Instructions ...
Страница 7: ...UltiMate 3000 Series SD RS and BM Pumps Operating Instructions Page V ...
Страница 8: ...UltiMate 3000 Series SD RS and BM Pumps Page VI Operating Instructions ...
Страница 14: ...UltiMate 3000 Series SD RS and BM Pumps Page vi Operating Instructions ...
Страница 24: ...UltiMate 3000 Series SD RS and BM Pumps Page 10 Operating Instructions ...
Страница 58: ...UltiMate 3000 Series SD RS and BM Pumps Page 44 Operating Instructions ...
Страница 74: ...UltiMate 3000 Series SD RS and BM Pumps Page 60 Operating Instructions ...
Страница 116: ...UltiMate 3000 Series SD RS and BM Pumps Page 102 Operating Instructions ...
Страница 136: ...UltiMate 3000 Series SD RS and BM Pumps Page 122 Operating Instructions ...
Страница 200: ...UltiMate 3000 Series SD RS and BM Pumps Page 186 Operating Instructions ...
Страница 204: ...UltiMate 3000 Series SD RS and BM Pumps Page 190 Operating Instructions ...
Страница 228: ...UltiMate 3000 Series SD RS and BM Pumps Page 214 Operating Instructions ...
Страница 236: ...UltiMate 3000 Series SD RS and BM Pumps Page 222 Operating Instructions ...