2
Choose the
Verification Method
.
Setting a
Confidence Level
is generally recommended. See also
.
If you enable
Absolute Bits/Errors
, the measurement proceeds with the next
amplitude as soon as one of the two numbers is reached.
3
If desired, set a
Relax Time
.
.
Before Running the Test
Before running the test, set up the instrument to ensure that the requirements of
the selected standard are met. Recommended test pattern and additional jitter are
displayed in the standard's description field on the
Standard
tab of the
Jitter
Tolerance Properties
dialog.
For details on the recommendations and their dependencies see
Description
in the
“Standard Specification” on page 462
.
For DUTs that require a training sequence it is essential to prevent clock loss at
the beginning or end of the test. To avoid clock loss, the sinusoidal jitter component
has to be enabled. Since the sinusoidal jitter for the test is generated internally,
the jitter amplitude must be set to zero:
1
Press
Jitter Setup
from the
Jitter
submenu.
8
Jitter Tolerance Tests
458
Agilent J-BERT N4903B High-Performance Serial BERT
Avoiding Clock Loss
Содержание J-BERT N4903B
Страница 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Страница 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 486: ... Input Timing Setup 2 Input Timing Setup 3 9 Solving Problems 486 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 487: ... Input Timing Setup 4 Input Timing Setup 5 Solving Problems 9 Agilent J BERT N4903B High Performance Serial BERT 487 ...
Страница 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...