Understanding the Output Protection Circuit
The generator module offers a huge flexibility for external termination schemes
and external termination voltages to address common technologies. For details,
please refer to the Technical Data Sheet.
An internal protection circuit continuously monitors the voltages of clock, data,
aux data and trigger output. It becomes active if the termination voltage is wrongly
adjusted.
The output protection circuit may become active if the output voltage Vout
becomes higher than Vhi+0.5V or lower than V
lo
-0.5V.
N O T E
The output protection circuit is shared among two output drivers each. Data Out
and Clock Out as well as Aux Data Out and Trigger/Ref Clock Out are supervised
by one circuit together. As a consequence, if an output voltage violation occurs,
both outputs are disabled together for a short period of time.
If an output voltage violation occurs, the output voltages Vhi, Vlo and Vterm are
re-programmed to safe values, typically to the externally measured termination
voltage. Afterwards the output drivers are re-enabled.
The GUI shows an error message, listing the effected outputs and highlights the
Outputs On
status indicator:
In the corresponding Output screen of the PG Setup, the output settings that
currently don’t match the electrical output settings are highlighted in yellow.
4
Setting up the Pattern Generator
124
Agilent J-BERT N4903B High-Performance Serial BERT
Содержание J-BERT N4903B
Страница 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Страница 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 486: ... Input Timing Setup 2 Input Timing Setup 3 9 Solving Problems 486 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 487: ... Input Timing Setup 4 Input Timing Setup 5 Solving Problems 9 Agilent J BERT N4903B High Performance Serial BERT 487 ...
Страница 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...