How Can You Tell if Your Synchronization is False?
You may suspect false synchronization under the following conditions:
• You are using a pattern other than PRBS and the error detector gains sync, but
it measures a constant, fixed error ratio.
• You are using a pattern other than PRBS and the error detector gains sync, but
auto-search functions (Auto Align, Clock/Data Center, 0/1 Threshold Center)
repeatedly fail.
In a false sync, the sync threshold BER of 1E-3 may be met, but eye edges at
BER 1E-3 (required by an auto-search function) may not be found. This is
because BERs less than 1E-3 do not exist within the data eye.
If you suspect a false sync, try re-synchronizing at a sync threshold BER lower
than the fixed error ratio. If sync is acquired without the problems listed above,
then your previous sync was false. Your current sync should be on an exact
pattern alignment.
N O T E
While auto-search functions are in progress, the sync threshold BER is changed to
the same value as the alignment BER threshold. If you are using these functions
and want to consistently re-synchronize at a lower sync threshold, you must set the
alignment BER threshold to the same value as the sync threshold BER.
5
Setting up the Error Detector
210
Agilent J-BERT N4903B High-Performance Serial BERT
Содержание J-BERT N4903B
Страница 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Страница 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 486: ... Input Timing Setup 2 Input Timing Setup 3 9 Solving Problems 486 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 487: ... Input Timing Setup 4 Input Timing Setup 5 Solving Problems 9 Agilent J BERT N4903B High Performance Serial BERT 487 ...
Страница 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...