The oscilloscope should show the following picture:
If you see only a straight line, ensure that the pattern generator is not disabled and
that it is set up to generate a pattern (e.g. PRBS).
Settings for Clock Multiplier Test
Following are the settings to test the output of the clock multiplier:
• Preset Instrument State
• Set Data Rate to 3GB/s
• Set Clock Offset to 0V
• Set Clock Out Amplitude to 400mV
• Set PG Trigger Out Offset to 0V
• Enable Clock Multiplier in the Config window
• Set the following for Clock Multiplier:
• Input Frequency 3GHz
• Termination ON at 0V
Check the Output of the N4916B (Clock Multiplier Part) Using a
Scope ( Clock Multiplier 1)
Following are the steps to test clock multiplier 1 using a scope:
Solving Problems
9
Agilent J-BERT N4903B High-Performance Serial BERT
507
Содержание J-BERT N4903B
Страница 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Страница 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 486: ... Input Timing Setup 2 Input Timing Setup 3 9 Solving Problems 486 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 487: ... Input Timing Setup 4 Input Timing Setup 5 Solving Problems 9 Agilent J BERT N4903B High Performance Serial BERT 487 ...
Страница 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...