N O T E
All measurements (except Jitter Measurement) in "8B/10B Symbol Comparison",
"Bit Comparison without PCIe3 SKPOS" and "Bit Comparison without USB3.1
SKPOS" modes, automatically switch to BER mode and work in Bit Recovery Mode.
These changes are reflected in GUI too. Jitter measurements also works in "8B/
10B Symbol Comparison" mode and uses calculated BER (cBER) to check against
Target BER.
In case of Jitter Measurement, the auto re-sync automatically switches to manual
sync mode and then restores it back when the measurement is finished/aborted.
Understanding the Error Free Receiving in 8B/10B Symbol
Comparison
When RX detects incoming data correctly, the pattern looped back has the “same”
content and number of symbols except for filler primitives. These can be removed
or inserted by DUT to compensate for the speed differences of the clock domains.
Furthermore, 8B/10B coding defines two different (bitwise complementary) 10B
symbols for the majority of 8bit words based on a running disparity.
It may therefore happen that the data looped back by the DUT is transmitted with
a different than the incoming disparity, depending on the DUT’s starting disparity.
Both effects represent desired operation of the DUT and the pattern looped back,
although bit-wise different in content and number of bits, is an indicator for error
free data detection of the RX-under test, so that consequently in order to report
this different bit stream as error free the ED has to recognize the 10B Data and
filler symbols (of both disparities) and to ignore the insertion or deletion of filler
symbols.
The following figure shows the functional block diagram of Option A02.
Setting up the Error Detector
5
Agilent J-BERT N4903B High-Performance Serial BERT
183
Содержание J-BERT N4903B
Страница 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Страница 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 486: ... Input Timing Setup 2 Input Timing Setup 3 9 Solving Problems 486 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 487: ... Input Timing Setup 4 Input Timing Setup 5 Solving Problems 9 Agilent J BERT N4903B High Performance Serial BERT 487 ...
Страница 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...