• Optimizing the View of the Results. (See
“How to Optimize the View of the
).
How to Prepare the Fast Eye Mask Measurement
To prepare a Fast Eye Mask measurement to test a shielded cable:
1
Disable the pattern generator outputs by pressing the
0V (Disable)
button in
the
PG Setup -> Data Output
screen.
2
Use a shielded cable to connect the pattern generator's Data Out port and the
error detector's Data In port.
3
Terminate all non-connected pattern generator output ports with 50 Ω.
4
Use a shielded cable to connect the pattern generator's Data Out port and the
error detector's Data In port.
5
Switch to the
Pattern
panel and press
Pattern Select
. Select an appropriate
pattern, for example a pure 2^15-1 PRBS, for both the pattern generator and
the error detector.
6
For the pattern generator setup, you need to specify the logic levels and the bit
rate. Select
ECL
levels and a clock speed of 1250MHz in this example. This
corresponds to a clock period of 800ps. See
“Setting up the Pattern Generator
for more information.
7
Set up the error detector so that the input range and the termination matches
the pattern generator's levels:
– Select an
Input Range
from -2V to 0V
– Set the
Data Termination
to -2V
– Set the
Alignment BER Threshold
to 1E-6.
– Set the
Clock Setup
to
Clock Data Recovery
to get the error detector's
clock from the incoming data stream.
8
Enable the pattern generator outputs by pressing the
0V (Disable)
button.
9
Press
Sync Now
and then
Auto Align
to find the optimum sampling point. Check
that the synchronization and the alignment were successful. None of the error
indicators should show red.
The resulting bit error rate should be zero.
How to Execute the Fast Eye Mask Measurement
To run the Fast Eye Mask measurement:
1
Switch to the
Analysis
panel and then press the
Fast Eye Measurement
icon.
2
Press the
Start
button to execute the measurement.
6
Advanced Analysis
310
Agilent J-BERT N4903B High-Performance Serial BERT
Содержание J-BERT N4903B
Страница 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Страница 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 486: ... Input Timing Setup 2 Input Timing Setup 3 9 Solving Problems 486 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 487: ... Input Timing Setup 4 Input Timing Setup 5 Solving Problems 9 Agilent J BERT N4903B High Performance Serial BERT 487 ...
Страница 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...