The Interference Channel provides a switchable signal path with Nelco 4000-6 FR4
characteristics. This path can be connected between the pattern generator's Data
Out port and the DUT. For instructions refer to
“Connecting the J20 Option” on
.
The Interference Channel provides also a sinewave generator and a resistive
combiner. This makes it possible to add sinusoidal interference (amplitude
modulation) to the data signal.
The Interference Channel is built up as shown in the following figure.
The figure above shows the connections for a single trace segment combined with
sinusoidal interference.
N O T E
Note that all these relays and traces exist twice to support single-ended and
differential operation.
When intersymbol interference is switched off, there is a minimum trace length
between P1 and P2.
ISI must be enabled if sinusoidal interference shall be added to the signal. In this
case, the minimum trace is longer (refer to the technical specifications).
Data can be input by connecting the pattern generator's Data Out port to either P1
or P2.
Jitter Tolerance Tests
8
Agilent J-BERT N4903B High-Performance Serial BERT
403
Содержание J-BERT N4903B
Страница 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Страница 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 486: ... Input Timing Setup 2 Input Timing Setup 3 9 Solving Problems 486 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 487: ... Input Timing Setup 4 Input Timing Setup 5 Solving Problems 9 Agilent J BERT N4903B High Performance Serial BERT 487 ...
Страница 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...