Cancel
Click this button to cancel the
Auto Align
,
0/1 Threshold Center
, or
Data Center
functions while they are in progress. The following parameters will be returned to
their previous value or status:
•
Auto Align
Canceled
–
Data Delay
and
0/1 Threshold
values returned to previous.
– Status of
Data Inverted
and
Avg. 0/1 Threshold
checkboxes returned to
previous.
•
0/1 Threshold Center
Canceled
–
0/1 Threshold
value returned to previous.
– Status of
Avg. 0/1 Threshold
checkbox returned to previous.
•
Data Center
Canceled
–
Data Delay
value returned to previous.
BER Threshold
In this list, select an alignment BER threshold that is appropriate for your
application.
The alignment BER threshold is the pre-defined threshold used by the
Data
Center
,
0/1 Threshold Center
, and
Auto Align
functions to define the edges of the
data input eye in the time and voltage axes. You may wish to change the threshold
for the following reasons:
• Choosing smaller alignment BER thresholds will cause the auto-search
functions to set more accurate sampling points. However, if the BER threshold
is set lower than the residual BER of the measurement, the auto-search
functions will fail. 1E-7 is the smallest BER threshold available.
• Running the auto-search functions and viewing eye results at different BER
thresholds will allow you to see the eye margin at specific BERs. You can also
construct basic eye contour information.
Avg. 0/1 Threshold
Activate this checkbox to make the error detector continuously set the 0/1
threshold level to the average DC voltage level of the incoming data signal.
N O T E
Manual 0/1 threshold changes are impossible while this function is selected.
Setting up the Error Detector
5
Agilent J-BERT N4903B High-Performance Serial BERT
201
Содержание J-BERT N4903B
Страница 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Страница 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 486: ... Input Timing Setup 2 Input Timing Setup 3 9 Solving Problems 486 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 487: ... Input Timing Setup 4 Input Timing Setup 5 Solving Problems 9 Agilent J BERT N4903B High Performance Serial BERT 487 ...
Страница 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...