If this is selected, RJ/DJ separation is not available.
•
BER Threshold
To calculate the parameters for the given BER threshold. This is the BER level
for which output timing numerical values (phase margin, skew, etc.) are
calculated. It is also the upper limit of the BER range for RJ/DJ separation.
The BER threshold influences some of the parameters of the DUT Output Timing
measurement. You can also drag and drop the horizontal BER threshold in the
graphical display to change this value.
•
Min BER for RJ/DJ Separation
Lower limit of the BER range for RJ/DJ separation.
•
Residual BER for Estimated Total Jitter
BER level for which the estimated total jitter is calculated. See
for details on how it is used.
You can select the number of
Decimal Places
to be displayed in the table.
Graph Tab
On the
Graph
tab, you can use the several options to optimize the graphical display
according to your needs.
Choose between
Unit Interval
and
Seconds
to select the timebase for the display's
x-axis.
Choose between
Logarithmic
and
Linear
to select the scale for the display's y-
axis.
For example, a DUT Output Timing measurement displayed on a linear scale may
look like this:
6
Advanced Analysis
258
Agilent J-BERT N4903B High-Performance Serial BERT
Table Number Format
Timing Units
Scale
Содержание J-BERT N4903B
Страница 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Страница 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 486: ... Input Timing Setup 2 Input Timing Setup 3 9 Solving Problems 486 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 487: ... Input Timing Setup 4 Input Timing Setup 5 Solving Problems 9 Agilent J BERT N4903B High Performance Serial BERT 487 ...
Страница 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...