• Error Location Capture cannot be used.
• The function
Error Detector Pattern tracks the Pattern Generator Pattern
is
disabled. This function is not re-enabled when the BRM is terminated.
• BRM does not work above 11.5 Gb/s when the CDR is enabled.
BRM indicators:
• The mode is indicated in the status panel and in the BER Bar.
• The mode is indicated in all screen dumps and log files.
• The Measurement User Interface shows when this mode is used.
Sampling Point Setup - Procedures
In most cases you will set up the sampling point automatically. Even if you wish
to make some manual adjustments, it is recommended to start with automatically
aligning the sampling point.
N O T E
Before adjusting the sampling point, make sure the pattern synchronization mode
is set up properly. See
“Pattern Synchronization - Procedures” on page 216
details.
How to Set the Optimum Sampling Point Automatically
1
Click
Sampling Point Setup
menu item from the
ED Setup
submenu.
2
Select an appropriate BER threshold from the
BER Threshold
list.
3
Press
Auto Align
to automatically set the optimum sampling point.
You can stop the search for the optimum sampling point at any time by pressing
Cancel
.
How to Adjust the Data Input Delay
1
Click
Sampling Point Setup
menu item from the
ED Setup
submenu.
2
Select the
Avg. 0/1 Threshold
checkbox to use a good start point on the vertical
voltage axis.
3
Click
Data Center
to move the sampling point horizontally towards the center
of the eye at the selected voltage. You may click
Cancel
at any time.
4
Now you can use the numeric keypad or front panel knob to manually adjust
the value in the
Data Delay
field.
5
Setting up the Error Detector
196
Agilent J-BERT N4903B High-Performance Serial BERT
Содержание J-BERT N4903B
Страница 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Страница 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 486: ... Input Timing Setup 2 Input Timing Setup 3 9 Solving Problems 486 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 487: ... Input Timing Setup 4 Input Timing Setup 5 Solving Problems 9 Agilent J BERT N4903B High Performance Serial BERT 487 ...
Страница 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...