With this method you can still place the sampling point anywhere in the clock
cycle to find the optimum sampling point, even at low frequencies.
Clock 310 Mbit/s
Covered Range
3.226 ns
However, if the bit rate falls below appr. 310 Mbit/s, even with this method
gaps occur in the range of possible sampling points.
Clock 150 Mbit/s
Covered Range
with Gaps
6.66 ns
Gap
Gap
Gap
Gap
Gap
As a result, you might not be able to find the optimum sampling point, if it falls
in one of the gaps.
Clock Setup - Procedures
To set the clock rate:
1
Press
Clock Setup
in the
ED Setup
menu.
2
Specify whether you are using an
External Clock Source
or
Clock Data
Recovery
(CDR).
3
If you are using CDR, you need to approximately specify the clock rate in the
Clock Rate
field (with an accuracy of 0.01%).
4
If you are measuring bursts, deactivate
Auto Threshold
and set the threshold
manually. You can also press the
Measure
button to measure the current DC
level at the CDR input and set the measured value as CDR threshold.
5
Clock rates that are frequently used can be saved as presets in the
Preset
list.
To select a clock rate from the
Preset
list, click on it. The clock rate appears in
the
Clock Rate
field.
6
Specify the Loop Bandwidth within the range of 100 KHz to 12MHz. Specify the
CDR Fine Adjust
and
Peaking
values. When using a
Preset
, this is only required
when you want to change these values.
7 Enable
Spread Spectrum; and adjust the deviation (%) and deviation type as
required.
5
Setting up the Error Detector
178
Agilent J-BERT N4903B High-Performance Serial BERT
Содержание J-BERT N4903B
Страница 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Страница 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 486: ... Input Timing Setup 2 Input Timing Setup 3 9 Solving Problems 486 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 487: ... Input Timing Setup 4 Input Timing Setup 5 Solving Problems 9 Agilent J BERT N4903B High Performance Serial BERT 487 ...
Страница 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...