Explanation of the Fast Total Jitter Measurement
The Fast Total Jitter measurement is an optimized method to determine the total
jitter for devices that generate a very low error density (BER well below 10
-10
).
To measure (not estimate) the total jitter for a device with a BER of 10
-12
with
conventional methods, one usually needs to compare more than 10
12
bits for each
sample point. To measure a full eye opening this way with appropriate timely
resolution takes time (maybe days or weeks, depending on the data rate), and the
probability of seeing one or no error in 10
12
bits is not higher than 37%.
If one would compare 10
13
bits for the same device, the probability of seeing ten
errors is even lower (12 %), but the probability of observing no error is almost zero.
The Fast Total Jitter measurement implements a method that reduces the
measurement time considerably and provides a higher accuracy. It is based on
statistical and probability calculations.
The method was presented at the DesignCon 2005 and is described in
"Total Jitter
Measurement at Low Probability Levels, using Optimized BERT Scan Method"
,
included as 5989-2933EN.pdf.
We will not go into the details but provide an introduction.
Quite often, we do not need to measure the exact BER, but can stop the
measurement if we are sure that the BER is above or below a threshold. In a jitter
tolerance test, for example, we need just to assure that the device under test
6
Advanced Analysis
246
Agilent J-BERT N4903B High-Performance Serial BERT
The Uncertainty Band
Содержание J-BERT N4903B
Страница 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Страница 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 486: ... Input Timing Setup 2 Input Timing Setup 3 9 Solving Problems 486 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 487: ... Input Timing Setup 4 Input Timing Setup 5 Solving Problems 9 Agilent J BERT N4903B High Performance Serial BERT 487 ...
Страница 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...