This function is useful for providing a "starting point" for sampling somewhere
within the data eye on the vertical voltage axis. It is recommended to use this prior
to clock/data alignments.
When used with the following patterns, this function may provide a 0/1 threshold
that requires no further adjustment:
• 2^n-1 PRBS patterns
• Patterns with 50% mark densities (large amplitudes and no excessive noise)
Bit Recovery Mode
Click this button to put the error detector in bit recovery mode (BRM). For details
on BRM see
“Understanding Bit Recovery Mode” on page 195
.
Then, click
Auto Align
to set the sampling point.
Sampling Point Display
This area of the
Sampling Point Setup
window graphically illustrates the
parameters of the current sampling point and data eye. Click anywhere in the
sampling point display to set the sampling point to that location.
A green area in the display indicates the voltage range specified in the
Input
Range
fields for non-differential modes (Normal and Complement). The Serial
BERT is searching this range to find the eye when calculating the optimum
sampling point.
Return to Results
Click this button to return the
0/1 Threshold
and
Data Delay
to the values
measured during the last auto alignment. The
Data Inverted
and
Avg. 0/1
Threshold
checkboxes will be returned to their previous state.
5
Setting up the Error Detector
202
Agilent J-BERT N4903B High-Performance Serial BERT
Содержание J-BERT N4903B
Страница 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Страница 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 486: ... Input Timing Setup 2 Input Timing Setup 3 9 Solving Problems 486 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 487: ... Input Timing Setup 4 Input Timing Setup 5 Solving Problems 9 Agilent J BERT N4903B High Performance Serial BERT 487 ...
Страница 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...