• Every window reduces the spectral power.
• Results obtained from different devices can only be compared if the same
window is used.
Spectral Jitter - Procedures
This section shows how to set up and use the Spectral Jitter measurement. As an
example, we measure the spectral jitter behavior of a shielded cable.
This requires the following steps:
•
“How to Prepare the Spectral Jitter Measurement” on page 351
•
“How to Execute the Spectral Jitter Measurement” on page 352
How to Prepare the Spectral Jitter Measurement
To use the Spectral Jitter measurement to test a shielded cable:
1
Disable the pattern generator outputs by pressing the
0V (Disable)
button in
the
PG Setup -> Data Output
screen.
2
Use a shielded cable to connect the pattern generator's Data Out port and the
error detector's Data In port.
3
Terminate all non-connected pattern generator output ports with 50 Ω.
4
Switch to the
Pattern
menu and press
Pattern Select
. Select an appropriate
pattern, for example a pure 2^15-1 PRBS, for both the pattern generator and
the error detector.
5
For the pattern generator setup, you need to specify the logic levels and the bit
rate. Select:
– Logic levels:
LVPECL
– Clock speed:
Gb Ethernet
(1.06250Gb/s)
“Setting up the Pattern Generator - Concepts ” on page 121
information.
6
Set up the error detector so that the input range and the termination matches
the pattern generator's levels. Select
Normal
as the
Active Input
, an
Input
Range
of 1-3V, and set the
Data Termination
to 1.3V.
7
Enable the pattern generator outputs by pressing the
0V (Disable)
button.
8
Also make sure that the error detector derives its clock from the incoming data
stream using Clock Data Recovery. See
“Clock Setup - Procedures” on page
Advanced Analysis
6
Agilent J-BERT N4903B High-Performance Serial BERT
351
Содержание J-BERT N4903B
Страница 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Страница 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 486: ... Input Timing Setup 2 Input Timing Setup 3 9 Solving Problems 486 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 487: ... Input Timing Setup 4 Input Timing Setup 5 Solving Problems 9 Agilent J BERT N4903B High Performance Serial BERT 487 ...
Страница 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...