Table 35
LED Name
Description
LAN
Constantly green when LAN is operable
Constantly red when LAN is not operable
Flashing green when device
identification is enabled
Constantly orange during boot phase of
the instrument
Power
Constantly orange when instrument is
OFF (in power standby mode)
Constantly green when instrument is ON
(powered)
Problems with the M8061A
Problems with the M8061A - Concepts
When you have opened the
External Instrument(s) - Config
window and have
selected
M8061A
from the external instruments, select
Mux with DeEmphasis
under
Enable Function
column in the list, the
M8061A Connection
dialog appears.
Cannot-Connect Problems
An error message may appear when you now press
Enable
.
Check the Power Supply
Ensure that the power cord is connected. Check for the steady green "STATUS"
light that appears on the ESM of the chassis indicating a power ready status of
the chassis which mounts the M8061A module. Check for the status LEDs present
9
Solving Problems
518
Agilent J-BERT N4903B High-Performance Serial BERT
Содержание J-BERT N4903B
Страница 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Страница 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 486: ... Input Timing Setup 2 Input Timing Setup 3 9 Solving Problems 486 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 487: ... Input Timing Setup 4 Input Timing Setup 5 Solving Problems 9 Agilent J BERT N4903B High Performance Serial BERT 487 ...
Страница 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...